Get Quotes
Download Full Catalogue
Model: CNME011310
Feature:
Energy sequences including adult mode and pediatric mode
Waveform: Biphasic Truncated exponential
The maximum energy up to 200J
User interface: voice prompts and LED prompts through the entire process
Particle and Water Ingress: IP55
Daily and monthly self-test
Optional PC software, to review the resuscitation incident information
Dual languages to switch and volume settings
Technical Specification:
Category | Specification |
Mode | Adult, Pediatric Adult mode(default mode): 150, 150, 200Joules; Pediatric mode: 50, 50, 75 Joules; |
Waveform | Biphasic truncated exponential |
Energy | 200-Joules max. |
Voice Prompt | Multiple languages to select |
Visual Indicators | LED indicators |
Charge time (New, at 25°C) | Less than 7 sec. to 150J Less than 3 sec. to 75J Less than 10 sec. to 200J Less than 3sec. to 50J |
Analysis Time | 9sec. |
Controls | three buttons- On/Off, Shock,set |
Accessories | |
Battery type | Non-rechargeable,LiMnO2 12V,2.8 Ah |
AED pads | Disposable Adult pads and disposable pediatric pads |
Environment Specification | |
Operating Temperature | 0°C to 50°C |
Operating Humidity | Relative humidity between 15% and 95% (non-condensing) |
Atmospheric pressure | 500hPa~1060hPa |
Storage and Transport Temperature | -20°C to 55°C |
Particle and Water Ingress | IP-55 (Battery installed) |
Drop and shock abuse tolerance | IEC60601-1-12 (1 meter, any edge, corner, or surface, in standby mode) |
Standard accessories:
Disposable Adult AED pad*1 ,
Carrying Bag*1 ,
Non-Rechargeable battery*1
Option accessories:
Disposable Pediatric AED pad
IrDA Adapter(Transfer the data stored )
AED Cabinet
Navigation