Model:CNME011309
Features:
1.Three-step defibrillation process.
2.Two-button operation.
3.Extensive voice and visual prompts for the operator.
4.Biphasic energy output.
5.Lock-out protection to prevent inadvertent defibrillation.
6.Continuous event recording for reporting each use to a printer or computer.
7.Weekly self-test to ensure readiness.
8.Color LCD screen
9.screen for ECG wave and settings.
10.Adult and Pediatric defibrillation.
11.Multi language voice and monitor.
12.SD Card for recording capacity.
13.Infra-red for data transfer.
14.Multi language.
15.LCD Battery Capacity Indicator.
16.CPR coaching (Optional).
17.ECG measurement by 3 lead cable (Optional).
Specifications:
Output | Biphasic Truncated exponential |
Energy Sequence(adult) | 150/150/200(Default) 150/200/200 200/200/200 Note: the Energy sequence can be changed using the computer IR port. |
Energy Sequence(Pediatric) | 50/70/85(Default) 70/70/85 70/85/85 85/85/85 Note: the Energy sequence can be changed using the computer IR port. |
Charge Time | 10 sec. to 200 J |
Analysis Time | 9 sec. |
The maximum time from the initiation of rhythm analysis to readiness for maximum Energy discharge after 15 shocks. | Less than 30 seconds |
The maximum time from initially switching power on to readiness for maximum Energy discharge after 15 shocks. | Less than 35seconds |
Audible Prompts | 19 audible prompts |
Visual Prompts | LED/LCD |
Controls | six buttons - On/Off, Shock, and 4 software-configurable buttons |
Output Energy Accuracy | ±15% into any impedance from 25 to 175Ω |
The Maximum Voltage | 1800±20V |
Output disabled when the PATIENT impedance is outside limits | 20Ω to 200Ω |
Recovery of the ECG input after defibrillation | <10s |