Get Quotes
Download Full Catalogue
Function Introduction:
The Air/Oxygen Blender is a medical device used to mix Medical Air and Oxygen into a gas source ranging from 21% to 100% oxygen. The capacity is to provide and convey accurately controlled mixed air and oxygen gas to the patient in clinical use. The device is not only used as open-ended oxygen deliverer, but also important to offer flexibility to adapt to many medical devices, such as infant incubator, Infant Radiant Warmer, Infant Oxygen Hood, Oxygen Therapy Instrument, Ventilator, CPAP, Resuscitator and Epidural Instrument, etc, especially suitable for neonatal intensive care units and ward oxygen delivery.(Ban for high blood oxygen patients)
Performance Characteristics:
Oxygen concentration level is 21%-100% continuous adjustable;
Oxygen concentration and flow rate is adjusted separately;
High precision regulating and continuous steady gas output;
Pneumatic control without power supply
Specification:
Oxygen concentration adjusting range and accuracy :
Oxygen concentration adjusting range: 21% -100% ;
Accuracy: ±5% 02 ( V/V ) .
Flow rate adjusting range: 0L/min -15L/min;
0L/min -30L/min (optional, without humidification bottle)
0L/min -30L/min (optional, without humidification bottle)
Input gas source pressure: 207kPa-517kPa
The Reverse Gas Flow of zero (The reverse gas flow caused by the gas flow from one gas source inlet to the other inlet port).
Packing dimension: 315*300*265mm; Gross weight 3.0kg
Technical Specification:
Model | Concentration valve | Flowmeter | Humidification bottle | Gas bleed valve | Main part dimension | Weight | Single/double patient use |
CNME-10B | single | Single | optional | Yes | 123*63*71cm | 0.84kg | Single patient use |
Navigation