Get Quotes
Download Full Catalogue
Model: CNME010422
Features:
Multiple ventilation modes, suitable for ICU, emergency department and operation room etc: (PRVC, APRV, DUOLEVEL, V SIMV, P SIMV, IPPV, A/C, PCV, PSV, SPONT/CPAP, SIGH, MANUAL).
Pressure and flow dual triggering mechanism make patient breathe better and more comfortable.
Rapid oxygen supply, automatically offer high flow rate oxygen within two minutes.
Pressure-Time, Flow-Time, Volume Time, ETCO2-Timeosillogram, real-time display the condition of patient in all respects.
Separate design of electronic circuit and gas flow rate keep safe running of ventilator.
Compact long life internal battery can provide emergency power, avoid risk of patient.
3 level alarming system display overall condition of machine, easier for users to find and solve problems according to indication.
Oscillogram freezing and recovery offer doctors a better understanding of the real-time situation for patient.
Self-check before operation, ensure every port of machine is in good condition.
Technical Specification:
Ventilation Mode | PRVC, APRV, DUOLEVEL, V-SIMV, P-SIMV, IPPV, A/C, PCV, PSV, SPONT/CPAP, SIGH, MANUAL |
Ventilator parameter | Tidal volume (Vt): 20 ~ 2000 mL Frequency (Freq): 1min~100min Oxygen concentration: 21%~100% I:E: 4:1~1:8 PEEP: 0cmH2O ~ 40 cmH2O Pressure limit: 20 cmH2O ~ 100 cmH2O |
Monitoring parameter | Frequency (Freq): 0/ min~ 100/ min Tidal volume(Vt): 20mL~2500mL MV: 0 L/min ~ 99 L/min Airway pressure: 0 cmH2O ~ 100 cmH2O Dynamic lung compliance monitoring: 1mL/cmH2O ~ 1000mL /cmH2O ETCO2 Concentration: 0mmHg ~ 152mmHg (0%-20%) Oxygen concentration: 15%~100% |
Oscillogram | P-T(Pressure-Time), F-T(Flow-Time), V-T(Volume-Time) ETCO2-T (End-Tidal CO2 -Time) P-V Loop(Pressure-Volume Loop), F-V Loop(Flow-Volume Loop), P-F Loop(Pressure- Flow Loop) |
Alarm and protection | The AC power failure alarm: Power failure or no connection Internal backup battery low voltage alarm≤11.3±0.3V No tidal volume: No tidal volume within 6 s High minute volume alarm: 5 L/min ~ 99 L/min Low minute volume alarm: 1 L/min~ 30 L/min High airway pressure alarm: 20 cmH2O ~ 100 cmH2O Low airway pressure alarm: 0 cmH2O ~ 20 cmH2O High oxygen concentration alarm: 19%~100% Low oxygen concentration alarm: 18%~99% Continuous pressure alarm: (PEEP+ 1.5 cmH2O)over16s Suffocation warning: 5~ 60s Fan error: Show on screen Oxygen deficit: Show on screen The maximum limited pressure: <12.5kPa |
Working condition | Gas source: O2, Air Pressure: 280 kPa- 600 kPa Voltage: 220V±22V Power frequency: 50Hz±1 Hz Input power: 900 VA (With air compressor) 250 VA (Without air compressor) |
Wooden case packing size | Main engine: 506*566*615 mm G.W..43 KG, NW.: 22 KG Air compressor: 683*687 * 1140 mm G.W. : 105 KG, N.W.:70 KG . |